Technical Specifications
 

Measuring range 0-100%
Accuracy +/-2% of full scale
Signal output 10-14mV
Linearity +/-2% of full scale
Response T90 13 sec
Temp Coefficient compensated
Shelf Life 6 months
Humidity Non-condensing 0-99% RH
Operating Temp 0 to 40℃
Recommended storage 0 to 25℃
Expected Life 60 months
Warrangty 18 months
Electrical Connector gold plated slip rings
 

 
 

Measurement Range: 0-100 % oxygen
Accuracy and Repeatability: < 1 % vol. O2when calibrated at 100 % O2
Zero Ofset: < 0.5 % vol. O2in 100% N2, applied 5 minutes
Linearity Error: < 3 % relative
Cross Interference: < 0.5 % vol. O2response to:
  10 % CO2balance N2
  80% N2O balance N2
  7.5% Halothane balance N2
  7.5 % Isofurane balance N2
  7.5 % Enfurane balance N2
  9% Sevofurane balance N2
  20% Desfurane balance N2
Infuence of Humidity: - 0.03 % rel. per % RH at 25C
Infuence of Pressure: proportional to change in oxygen partial
  pressure
Infuence of Mechanical Shock: < 1% relative after a fall from 1m
Operating Temperature: 0 to 50C
Temperature Compensation: built-in NTC compensation (see below)
Efect of Temperature between +5 C and +25 C: 3 % relative error
Compensation (steady state): between 0 C and +50 C: 8 % relative error
Operating Humidity: 0-99 % RH non-condensing
Long Term Output Drift: < 1 % vol oxygen per month
  typically < - 15 % relative over lifetime
Storage Temperature: -20 to +50 C
Recommended Storage: +5 to +15 C
Recommended Load: 10 kOhms
Warm-Up Time: < 30 minutes, after replacement of sensor
Weight: approximately 28 grams
  approximately 43 grams E61912 series

 